SPECIFICATIONS


Defibrillator

  • Output: Biphasic Truncated exponential
  • Energy Sequence: 150, 150, 200J
  • Charge Time: 8 sec. to150J
  • 10 sec. to 200 J(NEW battery)
  • Analysis Time: 9 sec.
  • The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery: Less than 30 seconds
  •  The maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks: Less than 35 seconds
  • The maximum time from the initiation of rhythm analysis to readiness for discharge: Less than 25 seconds
  • Audible Prompts: 20 audible prompts
  • Visual Prompts: LED prompts
  • Controls: Two buttons - On/Off, Shock
  • Output Energy Accuracy: ±15% into impedance from 25 to 75Ω
  • The Maximum Voltage: 1200±50V
  • Output disabled when the patient impedance is outside limits: 20Ω to 200 Ω

Battery

  • Part No: CR123A-4X2
  • Non-Rechargeable: 12V DC. 3.0 Ah
  • Capacity: 100 discharges at 200 Joules or 120 discharges at 150 Joules
  • Shelf Life (25℃±15℃): 10 years (5 years storage + 5 years standby)
  • 5 years standby (after installation)

Physical

  • Dimensions: 303 X 216 X 89 mm
  • Weight: 2 kg
  • Operating Temperature 0℃ to 40℃
  • Operating Humidity: Relative humidity between 30% and 95% (non-condensing)
  • Storage Temperature: -20℃ to 55℃ (without battery)
  • Storage Humidity: Up to 93% (non-condensing) (without battery)